Ten programmes
Select a cold-to-warm range suited to the expected film-forming temperature.
Product viewManufacturer images · Select a thumbnail to explore the instrument, platen and film-formation example.
Find the lowest temperature at which a water-based coating forms a continuous film. The MFFT 90 creates a controlled temperature gradient so you can assess film formation across a range of temperatures in one test.
Ask Novasys about samples, utilities, accessories and a quotation.
03 9752 3766Supplied and supported locally by Novasys Group.
Minimum film-forming temperature is the lowest temperature at which a dispersion coalesces uniformly as a thin film. Below that point, the dried material may fail to form a continuous film.
MFFT testing helps laboratories compare formulations, emulsions and coalescing-agent changes. Rhopoint’s MFFT 90 uses a temperature-gradient platen for tests based on ASTM D2354 and ISO 2115.
A defined test procedure helps you compare samples and identify the point where a continuous film forms.
Select a cold-to-warm range suited to the expected film-forming temperature.
A nickel-plated copper surface is cooled at one end and heated at the other.
Air or nitrogen passes over the film from the cold end to the warm end. The instrument includes air drying and flow control.
Use the on-screen operating guide and temperature readout to locate the transition after the sample dries.

Select the programme and allow the instrument to reach thermal equilibrium. Apply wet sample tracks using the cube applicator, then close the viewing cover.
When the material has dried, identify the film-formation boundary and use the cursor and display to read its temperature. Test duration depends on the material and conditions.
Featured model: MFFT 90 · 230 V
Minimum film-forming temperature testing.
| Operating temperature span | −10°C to 90°C across 10 programmes |
|---|---|
| Standards | ASTM D2354 and ISO 2115 |
| Platen (L × W) | 483 × 235 mm |
| Air requirement | 4 L/min at 100 psig |
| Cooling water | Mains pressure with gravity drain; specified inlet temperature below 15°C |
| Mains options | 220–240 V AC or 110–120 V AC |
| 230 V order code | A1212-002 |
| 115 V order code | A1212-002/115 |
| Dimensions (H × W × D) | 350 × 550 × 610 mm |
| Weight | 38 kg |
| Packed dimensions (H × W × D) | 590 × 800 × 690 mm |
| Packed weight | 53.7 kg |
Manufacturer specifications for MFFT 90. Confirm the selected voltage, programme, cooling-water temperature and laboratory connections with Novasys before ordering.
Each programme establishes a gradient between its cold and warm ends. The −10°C to 90°C specification describes the instrument’s overall span.
| Programme | Cold end | Warm end |
|---|---|---|
| 0 | -10°C | 8°C |
| 1 | -5°C | 13°C |
| 2 | 0°C | 18°C |
| 3 | 5°C | 23°C |
| 4 | 15°C | 33°C |
| 5 | 23°C | 50°C |
| 6 | 33°C | 60°C |
| 7 | 43°C | 70°C |
| 8 | 53°C | 80°C |
| 9 | 63°C | 90°C |
Share your material type, expected transition temperature and test procedure with Novasys.
Confirm your working temperatures and facilities when requesting a quotation.

Optional chiller shown. It is separate from the standard MFFT 90 kit.
Temperature programmes, specifications and instrument details.
Open datasheet ManualInstallation, operation and instrument care guidance.
Open manual ManufacturerFurther measurement, application and accessory information.
Open manufacturerSpeak with your local Novasys team about product selection and support.
MFFT is the lowest temperature at which a dispersion forms a uniform, continuous film when applied as a thin layer. It is relevant to water-based coatings, latex, emulsions and suitable adhesives.
No. MFFT 90 provides ten defined temperature gradients within that overall span. For example, programme 0 runs from −10°C to 8°C, while programme 9 runs from 63°C to 90°C. Select the range around your expected transition.
Rhopoint states that MFFT 90 conforms to ASTM D2354 and ISO 2115. Agree the sample preparation, film thickness and test conditions required by your procedure before comparing results.
The manufacturer specifies air at 4 L/min and 100 psig, plus mains-pressure cooling water with a gravity drain. It lists a water inlet temperature below 15°C; colder programmes can require colder water. Novasys can help review your installation and chiller needs.
Rhopoint distinguishes two-part adhesive cure assessment from MFFT. Two-part adhesives do not show an MFFT; their minimum cure temperature can be investigated by assessing the cured sample on the platen. Discuss the method and cleaning requirements with Novasys.
The standard kit lists water and air connectors, spare silica gel, manual and a 75 µm cube applicator. The chiller and additional applicators are optional. Use the enquiry pop-up and include your material, temperature range, utilities and voltage requirements.
Tell us what you test. We’ll help you plan the next step.